510(k) Data Aggregation

    K Number
    K141768

    Validate with FDA (Live)

    Manufacturer
    Date Cleared
    2014-08-04

    (34 days)

    Product Code
    Regulation Number
    892.1550
    Reference & Predicate Devices
    Predicate For
    AI/MLSaMDIVD (In Vitro Diagnostic)TherapeuticDiagnosticis PCCP AuthorizedThirdpartyExpeditedreview
    Intended Use

    The LOGIQ V5/LOGIQ V3 is general purposed ultrasound imaging and analysis systems providing digital acquisition, processing and display capability and clinical applications including: Abdominal, Obstetrical, Gynecological, Small parts, Vascular/Peripheral Vascular, Adult Cephalic, Pediatric, Musculoskeletal, Transcranial, Neonatal Cephalic, Transvaginal, Urological and Cardiac.

    LOGIQ V5/LOGIQ V3 is a general-purpose ultrasound system used for acquiring and storing ultrasound images for applications such as: Obstetrics; Abdominal, Obstetrical, Gynecological, Small parts, Vascular/Peripheral Vascular, Adult Cephalic, Pediatric, Musculoskeletal, Transcranial, Neonatal Cephalic, Urological and Cardiac. The system used for performing measurements, annotating, printing/reporting on the acquired/stored images.

    Device Description

    The LOGIQ V5/LOGIQ V3 is a general-purpose entry level ultrasound scanner from LOGIQ family for private clinics focusing on OB/GYN. The device is a general-purpose imaging and analysis system providing real-time digital acquisition, processing and display capability intended for general radiology imaging and evaluation with some cardiology and vascular applications.

    AI/ML Overview

    The provided document is a 510(k) premarket notification for the GE Healthcare LOGIQ V5/LOGIQ V3 Diagnostic Ultrasound System. It details the device's indications for use and compares it to predicate devices to establish substantial equivalence.

    Based on the provided text, the following information can be extracted regarding acceptance criteria and the study:

    1. A table of acceptance criteria and the reported device performance

    The document does not explicitly present a table of quantitative acceptance criteria for device performance. Instead, the "acceptance criteria" are implied through the comparison with predicate devices and compliance with relevant standards to demonstrate substantial equivalence. The reported "performance" is based on functional capabilities and adherence to safety and regulatory benchmarks rather than specific diagnostic accuracy metrics.

    Acceptance Criteria CategoryReported Device Performance/Compliance
    Intended UseSubstantially equivalent to predicate devices for diagnostic ultrasound imaging and fluid flow analysis. The LOGIQ V5/LOGIQ V3 has a subset of clinical indications available on the predicate LOGIQ F series (e.g., Fetal/Obstetrics, Abdominal/GYN and Urological, Small Organ, Pediatric, Neonatal Cephalic, Adult Cephalic, Cardiac, Peripheral Vascular, Musculo-skeletal Conventional and Superficial, Transcranial, Transvaginal, Contact Type: Surface, Cavitary, TEE).
    Imaging Capabilities/ModesSubstantially equivalent to predicate devices, having a subset of imaging modes (B; M; Color, Power, PW& CW Doppler modes, Color M-mode, Harmonic imaging, Combined modes) available on the predicate LOGIQ F series.
    TransducersSubstantially equivalent to predicate devices, having a subset of transducers (3Sc-RS, 4C-RS, E8C-RS, L6-12-RS, 8C-RS, RAB2-6-RS) available on the predicate LOGIQ F series. New indications (N) and previously cleared (P) indications are specified for each transducer model and clinical application (e.g., for 4C-RS: Fetal/OB-P, Abdominal-P, Small Organ-N, Peripheral Vascular-N).
    Processing & Display FeaturesSubstantially equivalent to predicate devices, including features like Image freeze, Multiple images, Pan/Zoom, Image maps (color & gray), Cine loop, Spatial & temporal filters, L-R / T-B image rev., Digital harmonics, TGC, Raw data access, LOGIQ View, CrossXBeam, SRI-HD, B-Flow, TVI, Anatomical-M mode (AMM), Auto Optimization.
    Safety and EffectivenessThe systems are manufactured with materials evaluated and found to be safe; acoustic power levels are below applicable FDA limits ("Track 3 (within FDA limits)"); designed in compliance with approved electrical and physical safety standards. Evaluated for acoustic output, biocompatibility, cleaning and disinfection effectiveness, thermal, electrical, electromagnetic, and mechanical safety. Conforms with AAMI/ANSI ES60601-1, IEC60601-1-2, IEC60601-2-37, NEMA UD 3, ISO10993-1, NEMA UD 2, ISO14971, and NEMA DICOM Set.
    Fundamental Scientific TechnologyEmploys the same fundamental scientific technology as its predicate devices.
    Measurement, Annotation, ReportingSimilar capability to predicate LOGIQ F series systems in terms of performing measurements, capturing digital images, reviewing and reporting studies.

    2. Sample size used for the test set and the data provenance (e.g. country of origin of the data, retrospective or prospective)

    The document explicitly states: "The subject of this premarket submission, LOGIQ V5/LOGIQ V5, did not require clinical studies to support substantial equivalence."

    Therefore, there is no clinical test set, no sample size, and no data provenance information for a clinical study to report. The evaluation relies on non-clinical tests and comparison to existing predicate devices.

    3. Number of experts used to establish the ground truth for the test set and the qualifications of those experts (e.g. radiologist with 10 years of experience)

    Since no clinical studies were required to support substantial equivalence, there was no test set requiring expert-established ground truth.

    4. Adjudication method (e.g. 2+1, 3+1, none) for the test set

    Not applicable, as no clinical test set was used.

    5. If a multi reader multi case (MRMC) comparative effectiveness study was done, If so, what was the effect size of how much human readers improve with AI vs without AI assistance

    Not applicable. The document describes an ultrasound system, not an AI or computer-assisted diagnostic device, and no clinical comparative effectiveness study was performed.

    6. If a standalone (i.e. algorithm only without human-in-the-loop performance) was done

    Not applicable. This is not an algorithm-only device; it is a general-purpose ultrasound imaging system.

    7. The type of ground truth used (expert concensus, pathology, outcomes data, etc)

    Not applicable, as no clinical test set requiring ground truth was used. The substantial equivalence relies on comparison of technical specifications, safety standards compliance, and functional capabilities with predicate devices, rather than diagnostic accuracy against a definitive ground truth.

    8. The sample size for the training set

    Not applicable. The device is an ultrasound imaging system. The document does not describe the use of machine learning or AI models with training sets. Its development involved standard quality assurance measures such as risk analysis, requirements reviews, design reviews, and various levels of testing (unit, integration, performance, safety, simulated use).

    9. How the ground truth for the training set was established

    Not applicable, as there is no training set described.